5) Connect the connector of X-RAY I/F cable to the power supply and joint the skin top spiral
and lock nut temporarily. Adjust the length of the cable so that the clearance of the skin top
spiral and insulation lock is 5mm. After adjusting the length, fasten the skin top spiral tightly.
Refer to the note in step 3.
6) Tighten the skin top spiral and lock nut using 15mm screw wrench.
7) After connecting the cables, attach the cover of power box and LAN cable connector guard.
*Different types of screw are used for backside and lateral side. Make sure to use proper
type of screw respectively on installation.
8) Connect the sensor cable, LAN cable, AC cable to the backside of the power box.
*1 Push the connector of the sensor cable into the connector terminal and fasten it tightly
until the lock ring stops.
*2 Only the AC cable attached to the product is allowed to use.
AC Power cable
Gap
[Improper connection]
2. Installation
Skin top spiral
Lock nut
Insulation lock
About 5mm clearances
Sensor cable
LAN cable
Lock ring
16
Connector
Fasten the sensor
cable while applying
the pressure
No gap
[Proper connection]