F9 Fetal & maternal Monitor Service Manual
10.4.10 Replacing the Bottom Interface Board
To remove the bottom interface board,
1) Separate the upper assembly and the lower assembly.
2) Disconnect all the wires on the board.
3) Remove the five screws securing the board.
To replace the bottom interfaces board,
Replace the bottom interfaces board in a reverse procedure. Refer to the following illustration
when connecting wires. Ensure they are all correctly connected.
(v) (vi)
(vii)
10.5 Disassembling the Upper assembly
10.5.1 Replacing the Main control board
To remove the main control board,
1) Separate the upper assembly and the lower assembly.
2) Disconnect all the wires on the main control board.
3) Remove the four screws securing the board.
(iv) (iii) ( ii) ( i )
- 56 -
Disassembling the Monitor
No. Connect to
i
Network Wire
ii
Power Control
Board
iii
Battery
iv
Signal Cable
v
FM Module
vi
Print Control Board
vii Power Board + Fan