F9 Fetal & maternal Monitor Service Manual
To replace the DECG module,
Replace the DECG module in a reverse procedure. Refer to the following illustration when
connecting wires. Ensure they are all correctly connected.
( i )
10.4.3 Replacing the Print Controlling Board
To remove the print controlling board,
1) Separate the upper assembly and the lower assembly.
2) Disconnect all the wires on the PCBA.
3) Remove the three screws securing the PCBA.
To replace the print controlling board,
Replace the print controlling board in a reverse procedure. Refer to the following illustration
when connecting wires. Ensure they are all correctly connected.
( i )
(iv)
(ii)
(iii)
(ii)
(iii)
(v)
- 50 -
Disassembling the Monitor
No. Connect to
i
Bottom Interfaces Board
ii
IUP Socket
iii
DECG Socket
No. Connect to
i
Paper Sensor
ii
Paper Drawer
Detector
iii
Bottom
Interfaces
Board
iv
Print head
(2 Sockets)
v
Gear Assembly