Disassembling the Monitor

9.4.8 Replacing the Transducer Socket Board (wired model only)

To remove the transducer socket board,
1) Separate the upper assembly and the lower assembly.
2) Remove the five screws on the transducer socket board.
3) Remove the rubber gaskets on the sockets.
To replace the transducer socket board,
Replace the transducer socket board in a reverse procedure.

9.4.9 Replacing the Bottom Interface Board

To remove the bottom interface board,
1) Separate the upper assembly and the lower assembly.
2) Remove all the cables on the bottom interface board.
3) Remove the four screws on the bottom interface board.
To replace the bottom interface board,