F9 Fetal & maternal Monitor Service Manual

Chapter 2 Overview

NOTE:
This service manual is written to cover the maximum configuration. Therefore, your
model may not have some of the parameters / modules described, depending on what
you have ordered.

2.1 Introduction

This service manual is a reference for periodic preventive maintenance and corrective service
procedures for the F9 fetal & maternal monitor.
It provides troubleshooting information, assembly procedures, instructions for functional testing
and performance verification. It is intended for use only by technically qualified service
personnel.
When performing a service procedure, follow the instructions exactly as presented in this
manual. Failure to do so might damage the monitor, invalidate the product warranty or
lead to serious personal injury.

2.2 General Information

F9 Fetal & maternal Monitor (hereinafter called F9) is designed to comply with the
international safety requirements IEC/EN 60601-1 for medical electrical equipment. It is
class I equipment.
The protective categories against electric shock of the patient connections are:
1) Ultrasound (FHR1, FHR2)
3) Remote Event Marker (FM)
This symbol indicates that the electric shock defending grade of this instrument is Type B.
1) IUP
This symbol indicates that the electric shock defending grade of this instrument is Type BF.
WARNING
2) External TOCO
2) NIBP
3) SpO
4) Fetal Stimulator (FS)
2
- 4 -
Overview