F9 Fetal & maternal Monitor Service Manual
3) Disconnect all the wires.
4) Remove the studs/screw securing the FM module.
To replace the FM module,
Replace the FM module in a reverse procedure. Refer to the following illustration when
connecting wires. Ensure they are all correctly connected.
( i )
(ii)
(iii)

10.4.2 Replacing the DECG Module

To remove the DECG module,
1) Separate the upper assembly and the lower assembly.
2) Disconnect all the wires.
3) Remove the four screws securing the DECG module.
(v)
(vi)
(vii)
(iv)
- 49 -
Disassembling the Monitor
No. Connect to
i
Toco Socket
ii
US2 Socket
iii
US1 Socket
iv
Mark Socket
v
Speaker
vi
Bottom Interfaces Board
vii
Power Module