FTS-3 Fetal Telemetry System Service Manual

10.4 Disassembling the Upper assembly

10.4.1 Replacing the Charging Interface Board

To remove the charging interface board,
(1) Separate the upper assembly and the lower assembly.
(2) Turn on the connector swich with both index fingertips and remove the adhesive tape.
(3) Disconnect the wire.
(4) Remove four screws securing the charging interface board.
To install the charging interface board,
Install the charging interface board in a reverse procedure. Refer to the following illustration
when connecting the wires. Ensure it is correctly connected.
No.
1
Figure 10- 9
1
Figure 10- 10
Connect to
Indicative Light board