F9 Fetal & maternal Monitor Service Manual
3) Compare the measurement results with the reference results in the list. If all the results are in
the allowance range. Switch on the monitor and measure the voltage to earth of the pins
listed below.
10 TP3
11 TP9
12 TP8
13 TP2
4) If any one of the results exceeds the reference range, the FM module defection is confirmed.
Replacement of the module is recommended.
Item
Pin
10
TP3
11
TP9
12
TP8
13
TP2
14
TP13
15
TP11
16
TP4
Modules' Malfunction Verification
Reference result
(0.45±20% )V
(0.9±20% )V
(0.9±20% )V
(0.45±20% )V
(3.30±20% )V
(3.30±20% )V
(0.40±20% )V
- 43 -