F9 Fetal & maternal Monitor Service Manual
Setting
Set the Net Version of F9 to PHILIPS. (Refer to 5.8 Changing Net Version)
Restart F9.
Validation
1) Switch F9 to Demo mode.
2) Observe the surveillance window of the OB TraceVue system. If data from F9 is transferred to
the system, the connection succeeds. Otherwise check the connection or try on another
interface of the system.
NOTE:
The identity number of this interface is preset. The device No. of F9 does not take effect
in this system.
Connecting to Philips OB TraceVue System
- 67 -