Chapter 9 Modules' Malfunction Verification; Verifying Malfunction Of The Main Control Board - EDAN F9 Service Manual

Fetal & maternal monitor
Hide thumbs Also See for F9:
Table of Contents

Advertisement

F9 Fetal & maternal Monitor Service Manual

Chapter 9 Modules' Malfunction Verification

When a module is suspected of malfunction, verify it using the method described in this chapter.
NOTE: For all the interfaces, the first pin is the one with a square solder pad.

9.1 Verifying Malfunction of the Main Control Board

To verify the main control board,
1) Open the main unit using the procedures described in section 10.3.
2) Switch on the monitor. (Powered by AC or battery.)
3) Measure the voltage to earth of the pins listed below, using a multimeter:
J8
1
J31
2
Item
Pin
J8's first pin
1
J31's first pin
2
TP1
3
TP2
4
C35's positive pole
5
C16's positive pole
6
J2's first pin
7
C59
8
C61
9
Main control board
Modules' Malfunction Verification
Reference result
VCC+5V (5.0±5%) V
VCC33(3.3±5%)V
+12V (12.0±5%) V
PVCC(3.3±5%)V
VDDARM (1.9±5%) V
VDDALIVE (1.9±5%) V
VBB33 (3.3±5%) V
VDD_RTC (1.8±5%) V
Li_BAT (≥2.75V)
3
- 40 -
TP1
TP2
4
5
6
7
C35
C16
J2

Hide quick links:

Advertisement

Table of Contents
loading