F9 Fetal & maternal Monitor Service Manual
Chapter 9 Modules' Malfunction Verification
When a module is suspected of malfunction, verify it using the method described in this chapter.
NOTE: For all the interfaces, the first pin is the one with a square solder pad.
9.1 Verifying Malfunction of the Main Control Board
To verify the main control board,
1) Open the main unit using the procedures described in section 10.3.
2) Switch on the monitor. (Powered by AC or battery.)
3) Measure the voltage to earth of the pins listed below, using a multimeter:
J8
○
1
J31
○
2
Item
Pin
○
J8's first pin
1
○
J31's first pin
2
○
TP1
3
○
TP2
4
○
C35's positive pole
5
○
C16's positive pole
6
○
J2's first pin
7
○
C59
8
○
C61
9
Main control board
Modules' Malfunction Verification
Reference result
VCC+5V (5.0±5%) V
VCC33(3.3±5%)V
+12V (12.0±5%) V
PVCC(3.3±5%)V
VDDARM (1.9±5%) V
VDDALIVE (1.9±5%) V
VBB33 (3.3±5%) V
VDD_RTC (1.8±5%) V
Li_BAT (≥2.75V)
○
3
- 40 -
TP1
TP2
○
4
○
5
○
6
○
7
C35
C16
J2