F9 Fetal & maternal Monitor Service Manual

8.6 Network Failures

Phenomenon
The monitor can not connect
to a network.

8.7 Alarms Failures

Phenomenon
No audible alarm is activated.
The alarm indicator stays off.
No audible alarm or visual
alarm is activated.

8.8 Technique Alarms

Phenomenon
Fetus EQUIP MALF
Possible Cause
The RS232 or Ethernet socket
defective.
Bad
connection
between
bottom interfaces board and main
control board.
Main control board defective.
Overlapped device no. in the
network.
Incorrect sever IP setting.
Possible Cause
The audible alarm is temporarily
disabled.
Loudspeaker or wire defective.
Alarm indicator defective.
Alarm indicator board defective.
Program defective.
Possible Cause
Bad
connection
between
module, bottom interfaces board
and main control board.
FM module defective.
Main control board defective.
- 38 -
Troubleshooting
Solution
Replace the RS232 or Ethernet
socket.
the
Check the connection.
Replace main control board.
Change
device
monitor.
Set server IP correctly.
Solution
Switch on the audible alarm.
Replace the loudspeaker or the
wire.
Replace alarm indicator.
Replace alarm indicator board.
Update the software.
Solution
FM
Check the connection.
Replace FM module.
Replace main control board.
no.
of
the