Quick Start Wizard Via The Handheld Programmer - Siemens SITRANS LR250 Instruction Manual

Hide thumbs Also See for SITRANS LR250:
Table of Contents

Advertisement

The current selection is highlighted.
If required, scroll to a new selection or key in a new
value and press Right ARROW
displays the new value.
Press Mode

Quick Start Wizard via the handheld programmer

Notes:
• The wizard is a complete package and the settings are inter-related.
• Because the settings are inter-related, the initial Quick Start parameter values are
not default values and do not necessarily reflect the current device configuration.
• Do not use the Quick Start wizard to modify individual parameters: see instead
Parameter Reference
1. Quick Start
a.
Point the programmer at the display (from a maximum distance of 600 mm [2 ft.]),
then press Right ARROW
b.
Press Right ARROW
view.
c.
Press Right ARROW
values and move directly to the next item.
d.
To change a setting, scroll to the desired item or key in a new value.
e.
After modifying a value, press Right ARROW
ARROW
f.
Quick Start settings take effect only after you select Yes to Apply changes in step 1.7.
1.1. Application Type
Options
1.2. Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Options
Use a setting just faster than the maximum filling or emptying rate (whichever is
greater). Slower settings provide higher accuracy: faster settings allow for more
level fluctuation
7ML19985JE01
to return to Measurement mode.
on page 41.
to activate PROGRAM mode and open menu level 1.
twice to navigate to menu item 1.1 and open parameter
to open Edit mode or Down ARROW
to move to the next item.
LIQUID LOW DK (low dielectric liquid)
SLOW
0.1 m/minute
MED
1.0 m/minute
FAST
10.0m/minute
SITRANS LR250 (HART) – INSTRUCTION MANUAL
to accept it. The LCD
to accept it and press Down
LIQUID
SOLID
APPLICATIO..
2.2.1.2
SOLID
LIQUID
LIQUID LOW DK
to accept default
Page 35