2.2.5.7. Range (Auto Suppression Range)
Defines the endpoint of the Learned TVT distance. Units are defined in sensor
units.
Values
a.
b.
c.
7ML19985JE01
Go to Input > TVT Setup > Range.
Determine the actual distance from the antenna reference point to
the material surface.
Subtract 0.42 m (16.5") from this distance, and enter the result.
Display before Auto False Echo Suppression
Example after Auto False Echo Suppression
false echo
false echo
SITRANS LR250 (HART) – INSTRUCTION MANUAL
Range: 0.00 to 20.00 m
Default: 1.00 m
Default
TVT
TVT
Hover Level
learned
TVT curve
Window
Echo
Marker
liquid
level
Page 75