(10) Press ENTER will perform 50A calibration. The system will set the output current to
50.00A automatically and the cursor will stop at position [4] as Figure 3-94 shows. Input
the current (4 digits after decimal point) read by Current Shunt (DVM) and press
ENTER
"
" to confirm.
S H O R T
O U T P U T
S E T T I N G O U T P U T
A C T U A L
S E T T I N G O U T P U T
A C T U A L
(11) The current calibration is done once the above actions are completed. Turn off the DC
Power Supply output and press "
3-95 shows, or press "
(12) Press "
EXIT
" to return to MAIN PAGE.
Improper shunt range selection may cause damage to the current shunt.
NOTICE
The calibration point may be different for other models (non 62012P-80-60), please operate it
following the instructions displayed.
[ C U R R E N T
T E R M I N A L
C U R R E N T
O U T P U T
C U R R E N T = 1 9 . 9 9 2 2 A
C U R R E N T
O U T P U T
C U R R E N T = 4 9 . 9 8 1 2 A
Figure 3-94
SAV E
" and "
" to return to Calibration page without saving.
EXIT
S A VE = Y E S
Figure 3-95
S E T T I N G C A L I B R A T I O N ]
A N D
P R E S S [ E N T E R ]
2 0 . 0 A
5 0 . 0 A
ENTER
"to save the calibrated data as Figure
Manual Operation
[
] 4
3-57