3.3.2.2 CURRENT LIMIT SETTING

1.
Use "
", "
V
L I M I T :
I
L I M I T :
V
S L E W
I
S L E W
T T L
V A L U E
< B I N A R Y
2.
Use the numeric (
Take example by 62012P-80-60, the output current range is 0~60A, but using this option
can narrow down its range by setting MIN and MAX.
When "
CURRENT
" is pressed to set the output current, the DC Power Supply allows setting
the current within the range of [MIN value ≤ user-defined value ≤ MAX value]. If the
setting exceeds the range, the BUZZER will beep (if BUZZER is set to ON) and the
main screen will prompt a warning message automatically as shown below. In the
meantime, the system will modify the current limit of MIN or MAX automatically.
I _ S E T = 1 5 . 0 0 A
V RAGNE = [V LIMIT MIN]V~[V LIMIT MAX]V
" keys to move the cursor to the column to be set.
M A X =
8 0 . 0 0 V
M A X =
1 5 . 0 0 A
R A T E
=
1 . 0 0 ( V / m S )
R A T E
=
1 . 0 0 ( A / m S )
=
=
0 0 0 0 0 0 0 0 >
~
) keys or "Rotary" (
0
9
1 2 . 0 0 A
Figure 3-18
[ O U T P U T
M I N =
0 0 0 . 0 0 V
M I N =
0 0 0 . 0 0 A
0 ( D E C )
Figure 3-19
) knob to set the value.
Manual Operation
C V
O N
9 6 0 . 0 W
S E T U P ]
3-13