5.4.2.
2-point adjustment
With a 2-point adjustment procedure the lower adjustment point must be between 0 and 40% of the measur-
ing range, and the upper adjustment point between 60 and 100% of the measuring range. If the adjustment
point is between 40 and 60% of the measuring range, automatically a 1-point adjustment procedure will be
executed instead.
possibility 1
possibility 2
possibility 3
lower adjustment point at 10% of measuring range
upper adjustment point at 90% of measuring range
characteristic line before adjustment
100%
characteristic line atfer adjustment
90%
80%
70%
60%
50%
40%
30%
20%
10%
10%
20%
30%
5.4.3.
Reset to factory settings
Customer-adjustment can be reset to the factory settings by checking the appropriate checkbox and subsequently clicking
the "reset" button.
lower adjustment point
0 - <40% of m.r.
40 - <50% of m.r.
0% of m.r.
40%
50%
60%
70%
80%
90%
100%
measuring range
upper adjustment point
60 - 100% of m.r.
100% of m.r.
50 - <60% of m.r.
29