FLOW METER for
COMPRESSED AIR and GASES
Manual
Hardware and Software
BA_EE772_e // V1.2 // technical data are subject to change // 193628

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EE772 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for E+E Elektronik EE772 Series

  • Page 1 EE772 Series FLOW METER for COMPRESSED AIR and GASES Manual Hardware and Software BA_EE772_e // V1.2 // technical data are subject to change // 193628...
  • Page 2 The content will be revised on a regular basis. These changes will be implemented in later versions. The described products can be improved and changed at any time without prior notice. © Copyright E+E Elektronik Ges.m.b.H. ® All rights reserved.
  • Page 3: Table Of Contents

    Table of contents - HARDWARE ......................1. GENERAL ....................1.1. Safety Instructions ....................1.1.1. Intended Use .................. 1.1.2. Installation, start up and control ....................1.2. Environmental aspects ..................2. PRODUCT DESCRIPTION ...................... 3. INSTALLATION ....................3.1. Mounting dimensions ..................3.2. Determining installation site ....................
  • Page 4: General

    Symbol Clarification This symbol indicates safety instructions. The safety instructions have to be carried out unconditionally. If disregarded loss, injury, or damage may be inflicted to people and property. In any case E+E Elektronik ® Ges.m.bH. cannot be hold responsible.
  • Page 5: Installation, Start Up And Control

    Consequential damages are excluded from the any liability. 1.2. Environmental aspects The products from E+E Elektronik are developed and designed in due consideration to the importance of the ® protection of the environment. Therefore, disposal of the product also should not lead to pollution of the environment.
  • Page 6: Product Description

    PRODUCT DESCRIPTION The flowmeter of the series EE772, based on the measurement principle of thermal mass flow, is suited for the measurement of flow of air and gasses in pipelines. Measurement of for instance the consumption of compressed air, nitrogen, helium, CO2 or other non-corrosive and non-flammable gasses. The EE772 measures the volume flow at standard conditions according to DIN 1343 (P =1023.25 mbar;...
  • Page 7: Installation

    INSTALLATION 3.1. Mounting dimensions EE772-A / EE772-B EE772-C Compact Remote probe (4.53) (5.71) The bottom part of the housing is mounted with 4 screws (not in the scope of supply). Max. screw diameter: 4.5 mm (0.17 inch). e.g. 4.2 x 38 mm DIN 7938H Screws cable gland M16x1.5 cable gland...
  • Page 8: Determining Installation Site

    3.2. Determining installation site • The installation site should be easy accessible and free of vibrations and shocks. • Observe at least 260 mm (10.24 inch) clearance above the housing with the signal conditioning, in order to be able to remove the sensing probe if necessary. •...
  • Page 9: Required Length Of Straight Pipe

    3.4. Required length of straight pipe The flowmeter should be installed as far as possible from disturbances of the flow. The causes for disturbance of the flow are for instance, reducers, elbows, T-pieces, valves, gate valves, etc. The specified measurement accuracy can be achieved only when the following straight inlet and outlet pipe lengths are installed: •...
  • Page 10: Assembly Of The Measurement Section

    3.5. Assembly of the measurement section In order to install or remove the measurement section the pipeline system should be depressurized. • All connections to be made with appropriated sealing material on the threads. • The sealing material should not change the area of the inner cross section of the pipe. It must be warranted that the connections after installation are free of leaks.
  • Page 11: Installation Of The Sensing Probe Into The Multicontroller

    3.6.2. Installation of the sensing probe into the MultiController Make sure that the pipe line is depressurized! • Remove transport protection cap of the head of the sensing probe. • Mount the sensing probe in the MultiController in such a way that the alignment pin fits in the recess on the MultiController.
  • Page 12: Installation Of The Sensing Probe Into The Hot Tap Valve

    3.6.3. Installation of the sensing probe into the hot tap valve The hot tap valve allows for the instalment and removal of the sensing probe under pressure, without inter- rupting the process flow. Operate the hot tap valve only with the hand, without tools! Functionallity: Position OPEN - MEASURE: The sensor head is immersed into the pipe.
  • Page 13: Electrical Connections

    ELECTRICAL CONNECTIONS Before electrical connections are made turn off the power supply first. If not observed the electronics can be damaged as a result. Only a qualified electrotechnical engineer may install the device. • Unscrew the four screws and remove the cover of the housing. •...
  • Page 14: Bus Output (Optional)

    4.2. Bus Output (optional) 4.2.1. M-Bus (Meter-Bus) The M-Bus (Meter-Bus) is a field bus for recording consumption data. The transmission is carried out serially on a reverse polarity protected two-wire line. The following measurements are transmitted: Standardized volumetric flow [Nm Mass flow [kg/h] Temperature [°C] Consumption reading [m...
  • Page 15: Control Components

    CONTROL COMPONENTS 5.1. Jumper J1 and J2 If the signal output is altered from relay to analogue output or vice versa, Jumper Output 1 has to be relocated. If the analogue output is altered from a current to a voltage output or vice versa, Jumper Out-1 has to be relocated.
  • Page 16: Display / Indicator With Keypad (Optional)

    5.3. Display / Indicator with keypad (optional) An optional two-line display is available for the flowmeter EE772. The display is an integral part of the cover of the housing and has two soft-keys for the control of the indicator. line 1 pushbutton - UP : 5.43 pushbutton - DOWN...
  • Page 17: Indication Of The Min / Max Values

    5.3.3. Indication of the MIN / MAX values Keep the DOWN key pressed for > 5 sec to indicate the MIN value. Keep the UP key pressed for > 5 sec to indicate the MAX value. MAX value press pushbutton UP >5s : 5.43 MIN value press pushbutton DOWN >5s 23.75 °C...
  • Page 18: Error Messages

    ERROR MESSAGES In case the flowmeter is equipped with the optional display, the following error message can be indicated. ERROR 01: sensing probe is not detected Cause: the sensing probe is not connect or is defect Effect: the display indicates “0” for all measurand. The analogue output defaults to the lowest configured value.
  • Page 19: Replacement Parts / Accessories

    REPLACEMENT PARTS / ACCESSORIES 8.1. Order Code Replacement Sensor EE772-R- Hardware Configuration Model compact ri-le direction od flow right to left compact le-ri direction od flow left to right remote probe Working range high (0,5...200 Nm/s) or (100...39370 ft/min) Measuring pipe - DN40 N040 diameter...
  • Page 20: Technical Data

    TECHNICAL DATA Measuring value Flow Measurand Volumetric flow at standard conditions acc. DIN 1343 = 1013.25 mbar = 0 °C (14.7 PSI) (32°F) Measuring range standardized volumetric flow in air DN40 2.26...904 Nm (1 1/2”) 1.33...531.8 SCFM DN50 3.50...1400 Nm (2”) 2.06...823.6 SCFM DN65...
  • Page 21: Factory Setting Of Outputs

    9.1. Factory setting of outputs SI-Unit Analogue output from unit [0...10 V / 0(4)...20 mA] standardized volumetric flow DN40 ≥DN50 1400 mass flow DN40 1140 kg/h ≥DN50 1780 kg/h standardized flow DN40 Nm/s DN50 Nm/s DN65 117,25 Nm/s DN80 77,41 Nm/s temperature °C...
  • Page 22: General

    CONFIGURATION SOFTWARE LIMITED LIABILITY E+E Elektronik shall not be held liable for any damages or consequential damages (for example, but not restricted to, loss of earnings, interruption of business, loss of information and data or any other financial losses) resulting from the installation, use or impossibility of use of an E+E Elektronik software product and any associated support services or non-performance of support services.
  • Page 23: Configuration Of The Usb Interface (Virtualcom)

    2.1. Configuration of the USB Interface (VirtualCOM) After the startup of the software, the correct VirtualCOM interface for the USB driver must be configured The number for the used USB interface can be found under: Start => Settings => Control Panel => System => Hardware => Device Manager The setting is done under menu “Extras”...
  • Page 24: User Interface

    User Interface Basic information: After retrieving the data from the transmitter, the basic information of the device is shown here. Status message: Here are the messages shown about the status and other information. Input screen: Input screen for the configuration or adjustment of the flowmeter. Menu tool bar: Selection of menu items.
  • Page 25: Transmitter

    4.2. Transmitter Read reads the actual configuration of the transmitter. Send uploads the ‘new’ configuration to the transmitter. The following settings are uploaded to the transmitter • Units • Output 1 • Output 2 • Display mode • Pressure transmitter Prior to uploading the ‘new’...
  • Page 26: Output Mode - Switch (Relay)

    5.1.5. Output mode – switch (relay) In the field for the “Switch-mode”, one can select “Hysteresis” or “Window”. The field for “Type” is to determine the switch action of the relay, NO = Normally Open (activate to close), NC = Normally Close (activate to open). Under “Measuring range”...
  • Page 27: Output Mode - Pulse

    5.1.6. Output mode – pulse If output 2 is configured for pulse, the measurand can be consumption only. Under “Pulse”, the duration of the pulse and the pulse value (Significance level of pulse) can be freely configured. Volume Flow [m Number of Pulses Pulse Value [m /Pulse]...
  • Page 28: Adjustment

    5.4. Adjustment The user can perform an adjustment for the measurands normflow and temperature in air. The configuration software distinguishes between a 1-point and a 2-point adjustment automatically, depending on how many reference points for adjustments are entered. The values entered for the customer’s adjustment are stored in the electronics of the sensing probe and are therefore not lost if the electronics of the signal conditioning are replaced (see page 6) If the checkbox “Performing customer-adjustment”...
  • Page 29: 2-Point Adjustment

    5.4.2. 2-point adjustment With a 2-point adjustment procedure the lower adjustment point must be between 0 and 40% of the measur- ing range, and the upper adjustment point between 60 and 100% of the measuring range. If the adjustment point is between 40 and 60% of the measuring range, automatically a 1-point adjustment procedure will be executed instead.
  • Page 30: Measuring Values Overview

    5.5. Measuring values overview The tab measuring values provides an overview of the retrieved actual measurement values of the flowme- ter (transmitter). Clicking on “Fetch values” will retrieve the actual measurement and MIN / MAX values for flow, volume flow, temperature, mass flow and pressure (only if a pressure transmitter is connected) from the transmitter –...
  • Page 31: Changing The Standard Conditions

    5.6.2. Changing the standard conditions The flowmeter is factory-set to standard conditions conforming to DIN 1343. Factory setting: P = 1013.25 mbar, t = 0°C (273.15 K) The corrected volume flow measured value is calculated in line with the standard conditions set. 5.6.3.
  • Page 32 HEAD OFFICE: E+E ELEKTRONIK Ges.m.b.H. Langwiesen 7 A-4209 Engerwitzdorf Austria Tel: +43 7235 605 0 Fax: +43 7235 605 8 info@epluse.com www.epluse.com SALES OFFICES: E+E CHINA / BEIJING Tel: +86 10 84992361 info@epluse.cn www.epluse.cn E+E CHINA / SHANGHAI Tel: +86 21 61176129 info@epluse.cn...