3.6.3.

Installation of the sensing probe into the hot tap valve

The hot tap valve allows for the instalment and removal of the sensing probe under pressure, without inter-
rupting the process flow.
Operate the hot tap valve only with the hand, without tools!
Functionallity:
Position OPEN - MEASURE: The sensor head is immersed into the pipe.
Position CLOSE - SERVICE: The hot tap valve pressure tight closed. You can remove the sensor probe.
OPEN-MEASURE
Turn the hot tap valve in position CLOSE - SERVICE and mount the sensing probe in the hot tap
valve in such a way that the alignment pin fits in the recess on the sealing cone.
Screw the retainer nut by hand so far that a certain resistance is noticeable.
Check the correct installation position of the flowmeter. The alignment pin must fit in the recess on
the sealing cone.
Tighten the retainer nut by hand. Tightening by hand should be sufficient. However, if the seal is
not leak tight carefully tighten the nut with an appropriate tool a bit further.
Turn the hot tap valve in position OPEN - MEASURE.
The mechanical installation of the flowmeter is therewith completed.