Changing The Standard Conditions; Pressure Compensation; External Pressure Transmitter For Pressure Compensation; Bus Configuration - E+E Elektronik EE772 Series Manual

Flow meter for compressed air and gases
Hide thumbs Also See for EE772 Series:
Table of Contents

Advertisement

5.6.2.

Changing the standard conditions

The flowmeter is factory-set to standard conditions conforming to DIN 1343.
Factory setting: P
= 1013.25 mbar, t
0
The corrected volume flow measured value is calculated in line with the standard conditions set.
5.6.3.

Pressure compensation

To achieve the highest measuring accuracy, the actual pressure can be entered in the field "pressure".
The "Send" button is used only to send the process pressure to the transmitter.
5.7.

External pressure transmitter for pressure compensation

In order to achieve the highest accuracy, the input from an external pressure transmitter will be very useful if
the pressure fluctuates strongly (e.g. 3 to 10 bar (45 to 150 psi)). An absolute pressure transmitter with a
2-wire loop powered 4 – 20 mA output should be used.
On the tab "Pressure transmitter" the measuring range can be entered.
5.8.

Bus configuration

If the flowmeter is equipped with an optional bus module, the data transfer rate and the network address can
be set on the "Bus configuration" tab.
The network address set is only used when the dip switches on the flowmeter PCB are set to 255 (see page 14).
= 0°C (273.15 K)