3.6.2.

Installation of the sensing probe into the MultiController

Make sure that the pipe line is depressurized!
Remove transport protection cap of the head of the sensing probe.
Mount the sensing probe in the MultiController in such a way that the alignment pin fits in the
recess on the MultiController.
Screw the retainer nut by hand so far that a certain resistance is noticeable.
Check the correct installation position of the flowmeter. The alignment pin must fit in the recess on
the ball valve.
Tighten the red coloured retainer nut by hand. Tightening by hand should be sufficient. However, if
the seal is not leak tight carefully tighten the nut with an appropriate tool a bit further
The mechanical installation of the flowmeter is therewith completed.