Installing And Removing The Filter Module (Fluorescent Option) - 3DHISTECH Pannoramic 250 FLASH User Manual

1.14 sp1
Hide thumbs Also See for Pannoramic 250 FLASH:
Table of Contents

Advertisement

Pannoramic 250 FLASH 1.14 SP1 User's Guide

5.3.5 Installing and Removing the Filter Module (Fluorescent Option)

1. Switch the HXP 120 illuminator off.
Crush hazard! Stop the running digitization process before you proceed
with filter module installation or removal.
2. Open the protective cover (see section 5.3.1 Opening the protective cover).
3. Select the desired filter in the corresponding software window (see section 4.5.5 Hardware
tab). The filter will be positioned in such a way that the user can access it without having to
turn the filter wheel manually.
4. Enter the respective parameters in the input fields.
5. Hold the filter module (3) by its handling pads (one on the right, one on the left (lateral)
side of the module) and insert it in forward tilted position from above into the upper spring-
loaded clamps (1) at the filter wheel.
Avoid touching the surface of the filters, and store the modules in their
own boxes when not used.
March 30, 2011
Figure 80 – Installing and removing the filter module
Warning!
Warning!
3DHISTECH Ltd.
5 Troubleshooting and Maintenance
120(130)

Advertisement

Table of Contents
loading

Table of Contents