Leakage
Allows for viewing some real-time measured values for leakage sensor diagnostics. If the leakage
sensor is not connected, the display will show MODE: SENSOR NOT CONNECTED and the
voltages VlKR and VlKS will be out of range below.
SETTING
DESCRIPTION
VLKR
Voltage at the reference sensor (approx. +0.65 V / +0.95 V).
VLKS
Voltage at the sample sensor (approx. +0.65 V / +0.95 V).
MODE
Mode of the sensor – OFF, STATE FLAG, WARNING, FATAL ERROR.

Service

The partial service menu provides access to the most common operations and is also compatible
with the service menu in older devices.
To access the partial service menu, press the MenU
hold the enTeR key for three seconds.

Figure 20

Partial Service Menu
The full service menu provides access to all available service operations.
To access the full service menu, press the MenU
arrow, up arrow, right arrow and down arrow buttons in sequence.
Figure 21
Full Service Menu
52
FRonT PAnEl conTRol
Leakage Screen
|
VERITY® 1741 UV-VIS DETEcToR
key,
select

SeRViCe

key,
select
SeRViCe
(7), and then press the left
(7), and then press and