NOTE
Connector blocks are meant for cables with size up to 1.5 mm
Diagnostics of the actual Io InTERFAcE status are accessed by pressing the MenU key and then
diagnostics
selecting
level and cloSE (l) means closed input, low level. For digital output is displayed oFF (for open
switch) and on (for closed switch).
The analog input, digital input 1 and 2, and digital output switches can be read through remote
control (ETHERnET).
diGiTAL inpUT (din1 And din2)
Function setting for inputs DIn1 and DIn2 on external interface:
SETTING
OFF
MAKE AUTOZERO
LAMP INIT
LAMP DEINIT
46
FRonT PAnEl conTRol
| interface. For digital inputs value oPEn (H) means opened input, high
DESCRIPTION
No function.
Pulse L causes the detector to be reset – autozeroed (AUTOZERO).
Pulse L causes initialization of the light sources (LAMP ON).
Pulse L causes deinitialization of the light sources (LAMP OFF).
|