Carbon Reference Material Run Conditions A - Micromeritics TriStar II 3020 Installation Instructions And Checklist

Surface area and porosity analyzer
Hide thumbs Also See for TriStar II 3020:
Table of Contents

Advertisement

A
C
NALYSIS
ONDITIONS
Field
Description
Isotherm Collection
Preparations
Free Space
P
and T
0
TriStar II 3020 Confirm Installation Instructions and Checklist
302-42870-04 (Rev A) — Sep 2017
F
S
ILE
ETTINGS
Enter
Reference material
Target Pressure
Do not adjust
Pressures
Options -
Use first pressure fixed dose: No
l
Use maximum volume increment: No
l
Absolute pressure tolerance: 5.0000 mmHg
l
Relative pressure tolerance: 5.0%
l
Fast evacuation: No
l
Evacuation time: 10.0mmHg/sec
l
Unrestricted evacuation from: 5.0 mmHg
l
Evacuation time: 0.30 hours
l
Leak test: No
l
Free space group: Measured
l
Lower dewar after free space: Yes
l
Evacuation time: 0.10 hours
l
Outgas test: No
l
P0 type: Measure
l
P0: 760.000 mmHg
l
Temperature: 77.30 K
l
Carbon Reference Material Run Conditions