A Run Conditions
D
C
EGAS
ONDITIONS
Leave these settings at the default. Do not modify.
A
C
NALYSIS
ONDITIONS
Field
Description
Isotherm Collection
Preparations
Free Space
P
and T
0
Equilibration
Backfill
A - 2
F
S
ILE
ETTINGS
F
S
ILE
ETTINGS
Enter
Run Conditions
Target Pressure
Pressures - Dp not adjust
Options -
Use first pressure fixed dose: No
l
Use maximum volume increment: No
l
Absolute pressure tolerance: 3.0000 mmHg
l
Relative pressure tolerance: 3.0%
l
Fast evacuation: Yes
l
Evacuation time: 0.30 hours
l
Leak test: No
l
Free space group: Measured
l
Lower dewar after free space: No
l
Evacuation time: 0.02 hours
l
Outgas test: No
l
P0 type: Measure
l
P0: 745.000 mmHg
l
Temperature: 77.30 K
l
Equilibration interval: 6 sec
l
Minimum equilibration delay at P/P0 > 0.995: 600 secs
l
Do not adjust
TriStar II 3020 Confirm Installation Instructions and Checklist
302-42870-04 (Rev A) — Sep 2017