...table continued from previous page
Specification Type
Internal System Battery Life
Typical Operating Time when System
is Operating at Intermediate Rate
Handling of Over-Infusion or
Under-Infusion
Bolus Volume at Release of Occlusion
(2 units per hour Basal)
Residual Insulin Remaining in the
Cartridge (unusable)
Minimum Audible Alarm Volume
Customer Technical Support: (877) 801-6901
Specification Details
4 years minimum under normal use conditions
During normal use, the intermediate rate is 2 units/hr; battery charge can be reasonably expected to last up to
7 days (5 days if using CGM) from a fully charged state to a totally discharged state
The method of delivery isolates the insulin chamber from the patient and the software performs frequent
monitoring of system status. Multiple software monitors provide redundant protection against unsafe
conditions.
Over-infusion is mitigated by continuous self-tests, layering of redundancies and confirmations, and numerous
other safeguard alarms. Users are required to review and confirm the details of all bolus deliveries, basal rates,
and temp rates to ensure certainty before initiating a delivery. In addition, once bolus deliveries are confirmed,
the user is given 5 seconds to cancel the delivery before it is started. An optional Auto-Off alarm triggers when
the user has not interacted with the pump's user interface for a pre-defined period of time.
Under-infusion is mitigated by occlusion detection and blood glucose monitoring as blood glucose entries are
recorded. Users are prompted to treat high blood glucose conditions with a correction bolus.
Less than 3 units with Unomedical Comfort
Approximately 15 units
45 dBA at 1 meter
Chapter 30 – Technical Specifications
(110cm) Infusion Set