Table 8-B. Sensor Stability Relative to YSI (Accuracy Over Time
Day
Number
of
Study
of Paired
2
Wear
CGM-YSI
1016
Original
Day 1
Software 505
740
810
Original
Day 4
Software 505
795
1096
Original
Day 7
Software 505
727
1. CGM readings are within 40–400 mg/dL, inclusive.
2. Both sets of study data are presented and are labeled as Original (SW10050) or Software 505 (SW10505).
Customer Technical Support: (877) 801-6901
) - (Pediatric, Ages 6-17 Years)
1
Mean
Median
Absolute
Absolute
Percent
Percent
Difference
Difference
21.2%
15.8%
12.7%
8.5%
16.0%
13.9%
8.1%
6.7%
15.1%
11.3%
10.4%
8.4%
Chapter 30 – Technical Specifications
Percent
Percent
Within
Within
15/15% YSI
20/20% YSI
48%
61%
75%
83%
52%
66%
89%
97%
63%
76%
80%
91%
Percent
Percent
Greater
Within
Than
30/30% YSI
40/40% YSI
78%
15%
91%
4%
87%
3%
100%
0%
89%
4%
98%
1%