Chapter 24 – Viewing CGM Data on Your t:slim X2 Pump
Sensor Session Ended
This symbol in the place where sensor glucose readings normally show means that your sensor session
has ended.
Wait 15 Minute Calibration Error
Either symbol will be shown in place of where the sensor glucose readings normally show means that the
sensor cannot calibrate right now. If you see this, enter at least one more calibration blood glucose value after
15 minutes. If the sensor still cannot calibrate after that, the sensor needs to be removed and a new sensor
needs to be inserted.
Failed Sensor
This symbol in the place where sensor glucose readings normally show means that the sensor has failed.
The sensor needs to be removed and a new sensor needs to be inserted.
Failed Transmitter
This symbol in the place where sensor glucose readings normally show means that the transmitter has
failed. The transmitter needs to be replaced.