t:slim X2 Pump Specifications
Specification Type
Classification
Size
Weight (with full disposable)
Operating Conditions
Storage Conditions
Atmospheric Pressure
Moisture Protection
Reservoir Volume
Cannula Fill Amount
Insulin Concentration
Alarm Type
table continued on next page ...
Customer Technical Support: (877) 801-6901
Specification Details
External PSU: Class II, Infusion Pump. Internally-powered equipment, Type BF applied part. The risk of ignition
of flammable anesthetics and explosive gases by the pump is remote. While this risk is remote, it is not
recommended to operate the t:slim X2 Pump in the presence of flammable anesthetics or explosive gases.
3.13˝ x 2.0˝ x 0.6˝ (L x W x H) - (7.95 cm x 5.08 cm x 1.52 cm)
3.95 ounces (112 grams)
Temperature: 41°F (5°C ) to 98.6°F (37°C)
Humidity: 20% to 90% RH non-condensing
Temperature: -4°F (-20°C ) to 140°F (60°C)
Humidity: 20% to 90% RH non-condensing
-1,300 feet to 10,000 feet
IPX7: Watertight to a depth of 3 feet for up to 30 minutes
3.0 mL or 300 units
0.1 to 1.0 units of insulin
U-100
Visual, audible, and vibratory