Table 11-B. Agreement to SMBG within CGM Glucose Ranges (Pediatric)
CGM Glucose
Study
2
Range
(mg/dL)
1
Original
Overall
Software 505
Original
40–60
Software 505
Original
61–80
Software 505
Original
81–180
Software 505
Original
181–300
Software 505
Original
301–350
Software 505
Original
351–400
Software 505
1. CGM readings are within 40–400 mg/dL, inclusive.
2. Both sets of study data are presented and are labeled as Original (SW10050) or Software 505 (SW10505).
Customer Technical Support: (877) 801-6901
Number
Percent within
of paired
15/15% SMBG
CGM-SMBG
16318
64%
4264
73%
487
44%
240
54%
1340
59%
399
64%
7084
62%
1650
72%
5627
69%
1526
79%
1176
65%
319
72%
604
58%
130
69%
Chapter 30 – Technical Specifications
Percent within
Percent within
20/20% SMBG
30/30% SMBG
76%
89%
84%
94%
55%
68%
71%
86%
70%
85%
76%
92%
74%
90%
84%
95%
80%
90%
89%
97%
77%
90%
83%
94%
72%
86%
79%
86%
Percent
Greater than
40/40% SMBG
5%
2%
19%
7%
7%
2%
5%
2%
5%
2%
4%
2%
6%
8%
329