Advertisement

Occlusion Alarm 2
What will I see on the Screen?
OCCLUSION ALARM
ALL DELIVERIES STOPPED!
Insulin delivery may be blocked.
Change your site and check your
BG in 1-2 hours.
CLOSE
Customer Technical Support: (877) 801-6901
What does it mean?
Your t:slim X2 Pump detected a second occlusion alarm shortly after the first occlu-
sion alarm and all deliveries have stopped.
How will the System notify me?
3 sequences of 3 notes or 3 vibrations depending on the volume/vibrate setting
selected in Pump Volume.
Will the System re-notify me?
Yes. The System will re-notify you every 3 minutes until you resume insulin delivery.
How should I respond?
Tap CLOSE. Change the cartridge, tubing, and infusion site to ensure proper delivery
of insulin. Resume insulin after changing the cartridge, tubing, and infusion site.
NOTE: Occlusion During Bolus
If the second occlusion alarm occurs during bolus delivery, after tapping CLOSE, a screen will appear
letting you know that the amount of bolus delivery could not be determined and was not added to your IOB.
Chapter 16 – t:slim X2 Pump Alarms
177