Chapter 30 – Technical Specifications
Rate of Delivery
25 Unit Bolus Delivery Speed
2.5 Unit Bolus Delivery Speed
20 Unit Prime
Bolus Duration
25 Unit Bolus Duration
2.5 Unit Bolus Duration
Time to Occlusion Alarm
Operating Rate
Bolus (3 units or Greater)
Basal (2 units/hr)
Basal (0.1 units/hr)
The time to occlusion alarm is based on insulin volume not delivered. During an occlusion event, boluses of less than 3 units may not trigger an occlusion alarm if no basal insulin
is being delivered. The bolus amount will reduce the time to occlusion depending on the basal rate.
290
2.97 Units/min Typical
1.43 Units/min Typical
9.88 Units/min Typical
8 minutes 26 seconds Typical
1 minute 45 seconds Typical
Typical
1 minute 2 seconds
1 Hour 4 Minutes
19 Hours 43 Minutes
Maximum
3 Minutes
2 Hours
36 Hours