Tandem T:Slim X2 User Manual page 311

Insulin pump
Hide thumbs Also See for T:Slim X2:
Table of Contents

Advertisement

Table 5-C. Hypoglycemic Alert and Detection Rate Evaluation in Reference to SMBG 30 Minutes Before and After
(Pediatric, Ages 2-5 Years)
Hypoglycemic
Alert Level
Study
(mg/dL)
Original
55
Software 505
Original
60
Software 505
Original
70
Software 505
Original
80
Software 505
Original
90
Software 505
Original
100
Software 505
1. Both sets of study data are presented and are labeled as Original (SW10050) or Software 505 (SW10505).
Customer Technical Support: (877) 801-6901
True
1
Alert Rate
3%
25%
11%
20%
29%
20%
35%
61%
51%
78%
64%
82%
Chapter 30 – Technical Specifications
False
Hypoglycemia
Alert Rate
Detection Rate
97%
57%
75%
100%
89%
62%
80%
100%
71%
77%
80%
100%
65%
85%
39%
100%
49%
89%
22%
100%
36%
91%
18%
100%
Hypoglycemia
Missed
Detection Rate
43%
0%
38%
0%
23%
0%
15%
0%
11%
0%
9%
0%
311