Table 6-C. Hyperglycemic Alert and Detection Rate Evaluation in Reference to SMBG 30 Minutes Before and After
(Pediatric, Ages 2-5 Years)
Hyperglycemic
Alert Level
Study
(mg/dL)
Original
120
Software 505
Original
140
Software 505
Original
180
Software 505
Original
200
Software 505
Original
220
Software 505
Original
240
Software 505
Original
300
Software 505
1. Both sets of study data are presented and are labeled as Original (SW10050) or Software 505 (SW10505).
Customer Technical Support: (877) 801-6901
True
1
Alert Rate
92%
97%
90%
98%
87%
99%
85%
98%
81%
100%
80%
99%
71%
95%
Chapter 30 – Technical Specifications
False
Hyperglycemia
Alert Rate
Detection Rate
8%
98%
3%
99%
10%
98%
2%
100%
13%
96%
1%
93%
15%
96%
2%
93%
19%
95%
0%
97%
20%
95%
1%
98%
29%
90%
5%
96%
Hyperglycemia
Missed
Detection Rate
2%
1%
2%
0%
4%
7%
4%
7%
5%
3%
5%
2%
10%
4%
315