Chapter 30 – Technical Specifications
...table continued from previous page
Specification Type
Basal Delivery Accuracy at all Flow
Rates (tested per IEC 60601-2-24)
Bolus Delivery Accuracy at all Volumes
(tested per IEC 60601-2-24)
Patient Protection from Air Infusion
Maximum Infusion Pressure Generated
and Occlusion Alarm Threshold
Frequency of Basal Delivery
Retention Time of Electronic Memory
when Internal System Battery is
Fully Discharged (including Alarm
Settings and Alarm History)
Infusion Set used for Testing
table continued on next page ...
278
Specification Details
±5%
The System is designed to vent automatically when there is a pressure difference between inside the
cartridge and the surrounding air. In certain conditions, such as a gradual elevation change of 1,000
feet, the System may not vent immediately and delivery accuracy can vary up to 15% until 3 units
have been delivered or elevation changes by more than 1,000 feet.
±5%
The pump provides subcutaneous delivery into interstitial tissue and does not deliver intravenous injections.
Clear tubing aids in detecting air.
30 PSI
5 minutes for all Basal Rates
Greater than 30 days
Unomedical Comfort
Infusion Set