MALUX SED OPH-810R User Manual page 25

Table of Contents

Advertisement

SED OPH-810R
User Manual
Attention: Please check the model of the charger before using it with this handheld. This handheld is
intended for use when supplied with power from the charger HKA00605008-3B.
Charging mode : Handheld using constant current and constant voltage charging mode.
If the battery is in the over-discharge state, enter the trickle charging stage. In normal times, First
entered the constant current charging stage, charging voltage is 5V and the charging current keep
the 800mA.
Entered the constant voltage charging stage when the battery voltage reaches a predetermined value,
the predetermined value needs to be calculated according to the battery calibration data. In the
constant voltage charging stage, the charging current decreases, when the charging current is
reduced to 90mA, the end of charging.
- 23 -