Troubleshooting Guide - Permobil X850 Corpus 3G User Manual

Hide thumbs Also See for X850 Corpus 3G:
Table of Contents

Advertisement

X850 Corpus 3G
6.1

Troubleshooting guide

The following troubleshooting guide describes a number of faults
and events which may occur when you use the wheelchair, together
with suggested remedies. Note that the guide cannot describe all the
problems and events which may occur and you should always
contact your service provider or Permobil in case of doubt.
Event
The wheelchair does not start.
The wheelchair cannot be driven.
The wheelchair switches itself off after a
certain period of inactivity (1 - 30 min).
148
Possible cause
Batteries discharged.
The cable connection to the control
panel has come loose.
Main circuit breaker switched to off
position after e.g. battery replacement.
Main circuit breaker tripped.
Battery charger connected.
Brake release activated.
Wheelchair locked.
The electronics' energy-saving mode has
been activated.
Troubleshooting
Remedy
Charge the batteries.
Insert the cable in the control panel.
Reset the main circuit breaker. See
page 142.
See page 142.
Stop charging. Disconnect the charging
cable from the wheelchair charger socket.
Reset the brake release.
Unlock the wheelchair.
Switch the wheelchair on again using the
start button on the control panel.