X850 Corpus 3G
WARNING!
Risk of fire or explosion - do not short-circuit
Exercise caution when using metallic objects during work with batteries. A
short-circuit can easily cause an explosion. Always use safety gloves and
goggles.
WARNING!
Risk for overheating, sparks or short-circuit
To interrupt the charging process, turn off the power supply at the switch on
the charger before removing the charging plug from the wheelchair's
charging socket. This is to avoid sparking and unnecessarily high wear on
the charging plug.
The charger may get hot and must not be covered. The charger must be
placed so that it has free space on all sides.
The charging jack must be replaced if it is damaged or gets hot during
charging. Both the jack on the charger cable and the wheelchair's
charging socket must be replaced if one part is damaged or worn. The
contacts must be replaced by qualified personnel.
114