Troubleshooting Guide - Permobil F3 Corpus User Manual

Hide thumbs Also See for F3 Corpus:
Table of Contents

Advertisement

F3 Corpus
8.1

Troubleshooting guide

The following troubleshooting guide describes a number of faults
and events which may occur when you use the wheelchair, together
with suggested remedies. Note that the guide cannot describe all the
problems and events which may occur and you should always
contact your service provider or Permobil should you have any
doubts.
Event
The wheelchair does not start.
The wheelchair cannot be driven.
The wheelchair switches itself off after a
certain period of inactivity (1 - 30 min).
260
Possible cause
Batteries discharged.
The cable connection to the control
panel has come loose.
Main circuit breaker switched to off
position after e.g. battery replacement.
Main circuit breaker tripped.
Battery charger connected.
Brake release activated.
Wheelchair locked.
The electronics' energy-saving mode has
been activated.
Troubleshooting
Remedy
Charge the batteries.
Insert the cable in the control panel.
Reset the main circuit breaker. See
page 246.
See page 246.
Stop charging. Disconnect the charging
cable from the wheelchair charger socket.
Reset the brake release.
Unlock the wheelchair.
Switch the wheelchair on again using the
start button on the control panel.