EDAN iSe Series User Manual page 92

Electrocardiograph
Hide thumbs Also See for iSe Series:
Table of Contents

Advertisement

iSE Series Electrocardiograph User Manual
Recommended separation distances between portable and mobile RF
communications equipment and the EQUIPMENT or SYSTEM
Recommended separation distances between
portable and mobile RF communications equipment and the iSE
The iSE is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the iSE can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the iSE as recommended
below, according to the maximum output power of the communications equipment.
Maximum
Separation distance according to frequency of transmitter (m)
output
power
rating
of
transmitter
P
2.3
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
150 kHz to 80 MHz
d
P
1.2
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d
P
1.2
0.12
0.38
1.2
3.8
12
- 79 -
EMC Information
800 MHz to 2.7 GHz
d
0.23
0.73
2.3
7.3
23

Hide quick links: