iSE Series Electrocardiograph User Manual
Electrodes
IEC
C1
White/Red
C2
White/Yellow
C3
White/Green
C4
White/Brown
C5
White/Black
Brown/Orange
C6
White/Violet
L
Yellow
R
Red
F
Locations
AHA
V1
Fourth intercostal space at the right border of the
sternum
Brown/Red
V2
Fourth intercostal space at the left border of the
sternum
Brown/Yellow
V3
Fifth rib between C2 and C4
Brown/Green
V4
Fifth intercostal space on the left midclavicular line
Brown/Blue
V5
Left anterior axillary line at the horizontal level of C4
V6
Left midaxillary line at the horizontal level of C4
Brown/Violet
LA
Left arm
Black
RA
Right arm
White
LL
Left leg
Preparing the Patient