EDAN iSe Series User Manual page 80

Electrocardiograph
Hide thumbs Also See for iSe Series:
Table of Contents

Advertisement

iSE Series Electrocardiograph User Manual
HR Range
Accuracy
Filter
DE12 / DE18 Performance
Leads Mode
Acquisition Mode
Analog-to-Digital
Converter
Time Constant
Frequency Response
Gain
Input Impedance
Input Circuit Current
Input Voltage Range
Calibration Voltage
DC Offset Voltage
Minimum Amplitude
Noise
CMRR
Pacemaker Detection (Single Channel Detection)
Amplitude
30 bpm – 300 bpm
1 bpm
AC Filter: 50Hz / 60Hz / Off
DFT Filter: 0.01Hz / 0.05Hz / 0.32Hz / 0.67Hz
EMG Filter: 25Hz / 35Hz / 45Hz / Off
Lowpass Filter: 350Hz / 300Hz / 270Hz / 150Hz / 100Hz /
75Hz
DE12: 12 standard leads
DE18: 18 standard leads
DE12: Simultaneously 12 leads
DE18: Simultaneously 18 leads
A/D: 24 bits
Resolution: 0.1192 uV/LSB
Sampling Frequency: 64,000 /sec/channel
≥ 5s
0.01 Hz – 350 Hz, -3 dB
1.25, 2.5, 5, 10, 20, 10/5, AGC (mm/mV) (±5%)
≥ 100 MΩ (10Hz)
≤ 0.01 μA
< ±5 mVp-p
1mV±2%
±900 mV, ±5%
20 μVp-p
≤ 12.5 μVp-p
≥140 dB (AC filter on)
≥123 dB (AC filter off)
DE12: ±500 uV to ±700 mV
- 70 -