iSE Series Electrocardiograph User Manual
The iSE is intended for use in the electromagnetic environment specified below. The
customer or the user of the iSE should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC/EN
61000-4-2
Electrical
fast
transient/burst
IEC/EN
61000-4-4
Surge
IEC/EN
61000-4-5
Power frequency
(50 Hz / 60 Hz)
magnetic field
IEC/EN
61000-4-8
IEC/EN 60601
test level
8 kV contact
15 kV air
2 kV for power
supply lines
1 kV line to line
2 kV line to ground
30 A/m
Compliance
level
8 kV contact
15 kV air
2 kV for power
supply lines
1 kV line to line
2
kV
line
ground
30 A/m
EMC Information
Electromagnetic
environment
guidance
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains
power
should be that of a
typical
commercial
hospital environment.
Mains
power
should be that of a
to
typical
commercial
hospital environment.
Power
frequency
magnetic fields should
be
at
characteristic of a typical
location
in
a
commercial or hospital
environment.
-
quality
or
quality
or
levels
typical