3. SERVICE MENU MAINTENANCE
Reset parameters: to perform a complete reset of the tests parameters (reset to factory
configuration).
CAN status: state of the internal network of the device.
Valve counter: to give an approximate state of the valves wear.
Device info: to know the information about the device, program version, built in components
etc.
Special cycles: to enable the user to adjust the pressures and service the valves and
pressure sensors. To make appear these special cycles in the special cycle menu, this
function must be activated.
4. RESULTS MENU
Results menu: to manage the test results.
Save on, to select the results saving mode, "NONE" = no save, "INTERNAL" to
save the results in the internal memory and "USB" to save the results in the USB
memory key connected to the device (in this mode, if no memory key is connected,
the results are lost.
Last results, to display the 6 lasts results carried out by the device and possibly to
delete them.
Statistics, to display the results statistics by program.
5. USB MENU
This is to save on an USB memory key the parameters to recover for later use, device
cloning, and program cloning or backup for device reconfiguration.
This is running if the parameters are unlocked.
MR-36300C-U
Quick start manual ATEQ F670 Page 21/25
Quick start - F670