ATEQ F670 Quick Start Manual
Hide thumbs Also See for F670:

Advertisement

Quick Links

QUICK START
ATEQ F670
Version 1.0
(Photo no contractual)
Reference: MR-36300C-U

Advertisement

Table of Contents
loading

Summary of Contents for ATEQ F670

  • Page 1 QUICK START ATEQ F670 Version 1.0 (Photo no contractual) Reference: MR-36300C-U...
  • Page 2 REVISIONS OF THE ATEQ F670 USER MANUAL Due to continuing improvements, the information contained in this user manual, the features and design of this device are subject to be changed without prior notice. Date Edition/Revision Reference Chapters up dating Week/Year...
  • Page 3: Table Of Contents

    Quick start manual TABLE OF CONTENTS TABLE OF CONTENTS ....................1 PREAMBULE........................2 1. DEFINITION OF THE ATEQ F670 ......................2 2. MEASUREMENT CHARACTERISTICS....................2 3. THE MAIN TYPES OF MEASUREMENT....................3 4. THE THREE TYPES OF TEST ......................3 5. MEASUREMENT CYCLE ........................3 INSTALLATION ......................
  • Page 4: Preambule

    Quick start - F670 PREAMBULE 1. DEFINITION OF THE ATEQ F670 The ATEQ F670 is a compact air/air TEST DIFFERENTIAL REFERENCE PART SENSOR PART leak detector used to test the air- tightness of parts. The method used is based on the measurement of a...
  • Page 5: The Main Types Of Measurement

    5. MEASUREMENT CYCLE WAIT FILL STABILIZATION TEST DUMP The measurement cycle consists of 5 phases: Coupling Stabilization Test Dump Cycle Start Fill time time time time time MR-36300C-U Quick start manual ATEQ F670 Page 3/25...
  • Page 6: Installation

    Quick start - F670 INSTALLATION 1. APPEARANCE OF THE ATEQ F670 1.1. F RONT FACE 1.2. R EAR SIDE MR-36300C-U Quick start manual ATEQ F670 Page 4/25...
  • Page 7: Electrics Connectors

    100 / 240 V AC OWER SUPPLY WITH FF SWITCH OPTION Supply the ATEQ F670 with the built in power supply, with a voltage between 100 and 240 V AC - 50 W. I: ON / O: OFF. 2.2. USB...
  • Page 8 2.5. J3 CONECTOR ANALOG OUTPUTS OPTION Connection for analog outputs. Pin 1: sensor 1 (plus). Pin 3: sensor 2 (plus). Pin 2: sensor 1 (minus). Pin 4: sensor 2 (minus). MR-36300C-U Quick start manual ATEQ F670 Page 6/25...
  • Page 9 200mA Max Output 5 End of cycle Output 5 End of cycle The compact mode is a software function which is activated in the CONFIGURATION/ AUTOMATISM / CHANGE I/O / OUTPUT menu. MR-36300C-U Quick start manual ATEQ F670 Page 7/25...
  • Page 10 (input 7) (input 8) (input 9) 17 à 32 33 à 64 65 à 128 With x who takes the 0 or 1 value in function of the program number to be called. MR-36300C-U Quick start manual ATEQ F670 Page 8/25...
  • Page 11 Input 4 (NPN or PNP)* Pin 14 Ground Analogue outputs Pin 15 Input 5 (NPN or PNP)* Pin 16 Ground * Inputs NPN or PNP following the strap position on the board. MR-36300C-U Quick start manual ATEQ F670 Page 9/25...
  • Page 12: Pneumatic Supply

    The air must be clean and dry. The supply pressure must always be between 4 and 8 bar (400 kPa and 800 kPa). MR-36300C-U Quick start manual ATEQ F670 Page 10/25...
  • Page 13: Pneumatic Connectors

    NEUMATICS TEST OUTPUTS These outputs enable parts to be connected (test, reference). The pressurization output is used for the addition of ATEQ accessories (Y valve). Inputs / Outputs on the rear side of the F670: Reference Output R Pressurization Output Test Output T 4.3.
  • Page 14: User Interfaces

    Scroll down or decrease numerical values Opening a menu, entering a parameter, confirmation of a parameter. Return to the previous menu or function, escape without modifying a parameter Programmable key by the user's preferences (see below). MR-36300C-U Quick start manual ATEQ F670 Page 12/25...
  • Page 15 Program defined: to access directly to the selected program parameters. Run program: to access directly to the current program parameterst (run program). Last results : to access directly to the test results menu. MR-36300C-U Quick start manual ATEQ F670 Page 13/25...
  • Page 16: Starting Up And Adjustments

    Access to the special cycle's service menu menu Access to the Access to the results menu, programs backup and display parameters Access to the USB Access to the connection configuration menu functions MR-36300C-U Quick start manual ATEQ F670 Page 14/25...
  • Page 17: Programs Management

    TEST TIME : 2.0 s ◄ value and validate with "OK". DUMP TIME : 0.0 s Press. UNIT: bar Max FILL 0.000 Do the same process for all the others parameters. MR-36300C-U Quick start manual ATEQ F670 Page 15/25...
  • Page 18: Parameters

    Note: if no extended function has been activated from the "More Functions" menus, the FUNCTION menu is empty. Edition, duplication, deletion or copy of a test program, program number to start: see the manuals CDROM for further information. MR-36300C-U Quick start manual ATEQ F670 Page 16/25...
  • Page 19: Start And Stop Measurement Cycle

    Press the RESET key to stop the current measurement cycle. The message "OK" show the device is pending a new test cycle. At the end of the measurement cycle, the test result is displayed. MR-36300C-U Quick start manual ATEQ F670 Page 17/25...
  • Page 20: Functions

    These functions, to appear in the program, must be activated first in the "More functions" menu. Then it must validate and configure it in function menu of the program. Each program can be individually personalized. MR-36300C-U Quick start manual ATEQ F670 Page 18/25...
  • Page 21: Others Menus

    ATR Learning: cycle to enter ATR parameters; this is to run at each switching on of the device or after a long time without measurement. Volume calculation: cycle to calculate the volume of the test circuit. MR-36300C-U Quick start manual ATEQ F670 Page 19/25...
  • Page 22: Menu Configuration

    Miscellaneous: Smart key, to configure the assigned to the "SMART KEY" key. Auto setup, to enables the automatic creation of a simple test program. MR-36300C-U Quick start manual ATEQ F670 Page 20/25...
  • Page 23: Service Menu Maintenance

    This is to save on an USB memory key the parameters to recover for later use, device cloning, and program cloning or backup for device reconfiguration. This is running if the parameters are unlocked. MR-36300C-U Quick start manual ATEQ F670 Page 21/25...
  • Page 24: Accessories And Characteristics

    1. ACCESSORIES PROVIDED 1.1. 100 230 V AC POWER SUPPLY The power supply cable of the F670 allows its connection to the mains supply network (from 100 to 240V AC). 2. ACCESSORIES IN OPTION Master leaks: the master leaks are used to check the device calibration.
  • Page 25: Technical Characteristic Of The F670

    100 à 240 V AC / 2 A Pneumatics connections: 3/5, 4/6 or 6/8 Weight (kg): about 8 Format : 19 inches (3U) +10°C to +45°C Running temperature: Storage temperature: 0°C to +60 °C MR-36300C-U Quick start manual ATEQ F670 Page 23/25...
  • Page 26: Errors And Faults

    If all the other checks do not resolve the problem, the instrument’s circuit may be checked. ATEQ does not accept any liability in regard to calibrations and settings to its instruments which are not carried out by its own personnel.
  • Page 27: Index

    Test Reject .........16 Measurement cycle ......3 Test time..........16 Measurements characteristics ....2 Touches de cycle........12 Minimum fill.........16 Type of test ...........3 More functions menu ......18 USB ............5 Navigation...........11 Navigation keys ........12 Operation fault ........24 MR-36300C-U Quick start manual ATEQ F670 Page 25/25...
  • Page 28 This document is the exclusive property of ATEQ. It may not be communicated, reproduced or used without prior consent.

Table of Contents