Section 8-3
Software Loading Procedure
8-3-1
General
This describes a full system software loading (Base Software Load Image + Arirang Application Load
Image) procedure.
Use this instruction when:
- Performing the HDD with new one (Base Software Load Image + Arirang Application Load Image)
- Performing the system revision-up (Application)
8-3-2
Parts Required
LOGIQ P6 Base Software DVD
LOGIQ P6 Application Software DVD
Blank CD-Rs (for patient image backup) or DVDs
Service Dongle
NOTE:
Patient Image Data will be deleted when all the drives are ghosted. Back up them on the CD-
R or DVD before starting software loading
NOTE:
Insert the service dongle into this service port located at the rear panel before installing the
Base System Software Load image
8-3-3
TimeRequired
Approximately 2hours
Page 8-104
Section 8-3 - Software Loading Procedure
LOGIQ™ P6/P6 P
S
RO
ERVICE
M
ANUAL