GE LOGIQ P6 Series Service Manual page 304

Table of Contents

Advertisement

GE H
EALTHCARE
5394152, Revision 5
D
IRECTION
8-2-32
RFC Assy
8-2-32-1
Tools
Common Phillips screwdrivers
8-2-32-2
Preparations
Shut down the system and switch off the main breaker.
8-2-32-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
2.) Remove the EMI Cover R. Refer to the
3.) Unscrew 2 screws (1,2) and eject the PCB.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-33
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-9-8
Section 4-3-4
Section 4-3-5
Section 10-5-2
8-2-32-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Page 8-78
Figure 8-118 RFC Assy
Functional Tests
Section
Functional Test / Diagnostic Test
RFC Assy Function Validation Procedure
System B/M-Mode Checks
System CFM and PWD Checks
Functional Checks (See Also Chapter 4)
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-14 "Right or Left Side Cover" on page
8-2-24 "EMI Cover L and R" on page
Power On/Boot Up
Power Off / Shutdown
LOGIQ™ P6/P6 P
S
RO
ERVICE
8-63.
Debrief Script
"Service Manual, Direction
5394152, Rev 1+, Section 8-2-32. Equipment
passed all required tests and is ready for use. "
M
ANUAL
8-48.