GE LOGIQ P6 Series Service Manual page 186

Table of Contents

Advertisement

GE H
EALTHCARE
5394152, Revision 5
D
IRECTION
Section 7-2
Gathering Trouble Data
7-2-1
Overview
There may be a time when it would be advantageous to capture trouble images and system data (logs)
for acquisition through remote diagnostics (InSite) or to be sent back to the manufacturer for analysis.
There are different options to acquire this data that would give different results.
7-2-2
Collect Vital System Information
The following information is necessary in order to properly analyze data or images being reported as a
malfunction or being returned to the manufacturer:
-
From the Utility>System>About screen:
Applications Software
-
-
System Image Software
-
-
Page 7-2
Product Name = LOGIQ™ P6/P6 Pro
Software Version
Software Part Number
Image Part Number
Image Date