GE LOGIQ P6 Series Service Manual page 290

Table of Contents

Advertisement

GE H
EALTHCARE
5394152, Revision 5
D
IRECTION
8-2-25
APS II Assy
NOTE:
APS Power Supply can not be adjusted.
8-2-25-1
Tools
Common Phillips screwdrivers
Nipper
8-2-25-2
Preparations
Shut down the system and switch off the main breaker.
8-2-25-3
Removal procedure
1.) Remove the Side Left Cover. Refer to the
2.) Remove the EMI Cover L. Refer to the
3.) Remove HDD power cable.
4.) Extract 3 power connector from back plane.
5.) Disconnect the AC input connector.
6.) Remove the Side Right Cover. Refer to the
7.) Remove the EMI Cover R. Refer to the
Page 8-64
AC Input Connector
Figure 8-103 APS II Connectors of backplane
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P6/P6 P
8-2-14 "Right or Left Side Cover" on page
8-2-24 "EMI Cover L and R" on page
8-2-14 "Right or Left Side Cover" on page
8-2-24 "EMI Cover L and R" on page
S
M
RO
ERVICE
ANUAL
8-48.
8-63.
8-48.
8-63.