Section 4-9Board Function Checks
4-9-1
PI128II Assy function check procedure
WARNING
WARNING
When Service Login for FRU Diagnostic, scan mode must be B-mode. After Diagnostic is completed,
the system should be rebooted
WARNING
WARNING
Diagnostics should be performed with probes which have 128 or more elements
1.) Turn on the system.
2.) After completion of echoloader loading, connect probe on port 1. Refer to the
a probe" on page
3.) Log into "Service Login". Refer to the
4.) Click "Diagnostics" on top menu.
5.) On left folder window, select 'LOGIQ P6 Diagnostics' => 'FRU test' => 'PI128 II'.
6.) Select 'PI128 II Overall' and execute test.
7.) All diagnostic test items shall be passed. Refer to the figure below.
NOTE:
For LP6 PRO, select 'LOGIQ P6 PRO Diagnostics' => 'FRU test' => 'P3RLY' and execute test.
8.) Shut down the system.
9.) Restart the system.
4-10.
5-7-2 "Global Service User Interface (GSUI)" on page
Figure 4-35 PI128II diagnostic test